Pseudo-polar drive patterns for brain electrical impedance tomography
Related Experiment Video
Updated: Jul 19, 2026

Electromagnetic Source Imaging in Presurgical Evaluation of Children with Drug-Resistant Epilepsy
Published on: September 20, 2024
Xuetao Shi1, Xiuzhen Dong, Wanjun Shuai
1Medical Electronic Engineering Department, Fourth Military Medical University, Xi'an, 710033, People's Republic of China.
Abstract:
Brain electrical impedance tomography (EIT) is a difficult task as brain tissues are enclosed by the skull of high resistance and cerebrospinal fluid (CSF) of low resistance, which makes internal resistivity information more difficult to extract. In order to seek a single source drive pattern that is more suitable for brain EIT, we built a more realistic experimental setting that simulates a head with the resistivity of the scalp, skull, CSF and brain, and compared the performance of adjacent, cross, polar and pseudo-polar drive patterns in terms of the boundary voltage dynamic range, independent measurement number, total boundary voltage changes and anti-noise performance based on it. The results demonstrate that the pseudo-polar drive pattern is optimal in all the aspects except for the dynamic range. The polar and cross drive patterns come next, and the adjacent drive pattern is the worst. Therefore, the pseudo-polar drive pattern should be chosen for brain EIT.