Construction and calibration of an instrument for three-dimensional ion imaging
Konrad Koszinowski1, Noah T Goldberg, Andrew E Pomerantz
1Department of Chemistry, Stanford University, Stanford, California 94305-5080, USA.
Abstract:
We describe a new instrument based on a delay-line detector for imaging the complete three-dimensional velocity distribution of photoionized products from photoinitiated reactions. Doppler-free [2+1] resonantly enhanced multiphoton ionization (REMPI) of H and D atoms formed upon photolysis of HBr and DBr in the range 203 nm < or = lambda photolysis < or = 243 nm yields radial speeds measured to be accurate within 1% of those calculated. The relative speed resolution is about 5% and limited by photoionization recoil broadening. A relative speed resolution of 3.4% is obtained for [3+1] REMPI, which minimizes the ionization recoil. We also determine the branching ratio between ground-state and spin-orbit-excited product channels and their associated anisotropies. We find that DBr photolysis dynamics differs slightly from its HBr counterpart.
More Related Videos
Related Concept Videos
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.
Mass Analyzers: Common Types
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview