Related Experiment Video
Updated: Jul 19, 2026

Closure of a Patent Foramen Ovale (PFO): An Intervention Sequence
Published on: December 23, 2022
Late migration of a Sideris buttoned device for occlusion of atrial septal defect
Timothy Cotts1, Peter J Strouse, Joseph N Graziano
1Division of Pediatric Cardiology, Department of Pediatrics, C.S. Mott Children's Hospital, University of Michigan, Ann Arbor, Michigan, USA. cottstim@umich.edu
Abstract:
We describe a 17-year-old patient with an atrial septal defect who underwent device closure with a second generation Sideris buttoned device at 4 years of age. She presented 13 years after the procedure with the acute onset of chest discomfort, at which time a chest radiograph showed migration of the wire of the right atrial counter-occluder to the lateral aspect of the right atrial wall, with an associated right atrial perforation. The patient underwent uneventful surgical device retrieval and atrial septal defect closure. Late migration of a portion of the Sideris buttoned device can occur, suggesting the need for continued follow up, and a high index of suspicion for device failure should a patient become symptomatic.
More Related Videos
Related Concept Videos
Mitral Stenosis III: Medical Management
Mitral Stenosis I: Introduction
Mitral Regurgitation I: Introduction
Mitral Regurgitation III: Medical Management