Related Experiment Video
Updated: Jul 19, 2026

07:28
An Efficient Method for Selective Desalination of Radioactive Iodine Anions by Using Gold Nanoparticles-Embedded Membrane Filter
Published on: July 13, 2018
A cost-effective modified micromethod for measuring urine iodine
Hussain Husniza1, Wan Mohamud Wan Nazaimoon
1Cardiovascular, Diabetes & Nutrition Research Centre, Institute for Medical Research (IMR), Kuala Lumpur, Malaysia.
Tropical Biomedicine
|October 17, 2006
Summary
A new urine iodine micromethod was developed and validated, showing high accuracy and precision. This method offers a reliable tool for assessing iodine status in populations.
Area of Science:
- Biochemistry
- Analytical Chemistry
- Public Health
Background:
- Accurate measurement of urinary iodine is crucial for assessing iodine status and monitoring public health interventions.
- Existing methods may require large sample volumes or specialized equipment, limiting widespread application.
Purpose of the Study:
- To establish and validate a modified micromethod for measuring urine iodine.
- To assess the accuracy, precision, and reliability of the developed micromethod.
Main Methods:
- A modified micromethod for urine iodine measurement was established.
- Validation involved comparison with a World Health Organization (WHO) reference method.
- Assessed parameters included correlation, agreement, sensitivity, recovery, linearity, and coefficients of variation (CVs).
Main Results:
- The micromethod demonstrated good correlation (r=0.962) and agreement with the WHO reference method.
- Sensitivity was determined to be 13.809 ug/L.
- Mean recoveries ranged from 103-114% across tested concentrations, with intra- and inter-assay CVs generally below 15%.
Conclusions:
- The modified micromethod is a valid and reliable tool for measuring urine iodine.
- Its accuracy, precision, and sensitivity support its use in public health programs for iodine status assessment.
Related Concept Videos
Redox Titration: Iodimetry and Iodometry
Iodometry and iodimetry are analytical methods used to determine the concentration of oxidizing or reducing agents using iodine. In iodometric titrations, the oxidizing analyte solution is usually acidified and treated with an excess of iodide ions, which generates an equivalent amount of iodine in equilibrium with triiodide. The released iodine is subsequently titrated directly against a standardized reducing agent. As the dilute iodine color becomes pale yellow, a few drops of freshly...
Urine Studies I: Urinalysis
Urinalysis is a widely used diagnostic test that analyzes urine's physical, chemical, and microscopic characteristics. Healthcare providers use it to detect and monitor various health conditions, including renal disease, urinary tract infections (UTIs), diabetes, and metabolic or systemic disorders.Components of UrinalysisUrinalysis consists of three primary components: physical, chemical, and microscopic examination. Each provides unique insights into the urine sample and, by extension, the...