A method detection limit for potential in vivo arsenic measurements with a 50 W x-ray tube
Ryan C N Studinski1, Fiona E McNeill, Joanne M O'Meara
1Department of Medical Physics, The Ottawa Hospital Regional Cancer Centre, Ottawa, Ontario K1H 8L6, Canada. rstudinski@ottawahospital.on.ca
Abstract:
Using a 50 W x-ray tube as the fluorescing source, an x-ray fluorescence system was designed to measure arsenic in superficial layers of tissue-simulating phantoms. The detection limit for arsenic in the phantoms at a setting of 35 kV and with 200 microm of a Mo filter was 0.40 +/- 0.06 microg As g(-1). This measurement results in an effective dose of 0.6 microSv.