Single-shot dual-energy subtraction mammography with electronic spectrum splitting: feasibility
Hans Bornefalk1, John M Lewin, Mats Danielsson
1Department of Physics, KTH, SE-100 44 Stockholm, Sweden. bornefalk@particle.kth.se
Abstract:
We present a single-shot dual-energy subtraction mammography technique using an energy sensitive photon counting detector. An electronic threshold near the middle of the X-ray spectrum discriminates between high- and low-energy photons, and allows the simultaneous acquisition of high- and low-energy images which can be combined to suppress anatomical clutter. By setting the electronic threshold close to 33.2 keV (the k-edge of iodine) the system is optimized for dual-energy contrast-enhanced imaging of breast tumors. This method eliminates the need for separate exposures which might otherwise lead to motion artifacts. The method is illustrated in phantom images.