Inter-session reproducibility of peak expiratory flow with standardized expiratory maneuvers
Charalampos Mandros1, Aris Tsiakalos, George E Tzelepis
1Department of Pathophysiology, University of Athens Medical School, Greece.
Background:
In adults performing forceful expiratory maneuvers, the length of post-inspiratory pause prior to forced expiration may influence the subsequently measured peak expiratory flow (PEF) and increase its variability. We investigated the effects of two different lengths of breath-hold at total lung capacity (TLC) on the short-term reproducibility of PEF in healthy volunteers.
Methods:
Forty-six healthy volunteers (age 34.6+/-8.5; 23 men) performed a series of maximal forceful expirations in two different test sessions, separated by approximately 2 weeks. In each test-session, PEF was measured with two different types of maneuvers. One maneuver (P) included a brief (<2s) post-inspiratory pause at TLC prior to forced expiration, whereas the second maneuver (NP) included no pause at TLC. The speed of inspiration to TLC was fast and similar for both maneuvers. In a given test session, all volunteers performed four efforts for each type of maneuver. The highest PEF for each maneuver was used for analysis. The Bland-Altman statistical analysis was used to determine inter-session reproducibility of PEF.
Results:
Within-maneuver analysis of the between-test session reproducibility of PEF showed that neither maneuver systematically biased the measured PEF (mean difference 0.02L/s for the P and 0.17L/s for the NP maneuver). Inter-maneuver between-test session analysis similarly showed that neither maneuver introduced a systematic bias in the maximal PEF (mean difference ranged from -0.15 to -0.01L/s). The limits of agreement were comparable in all maneuver-pair analyses.
Conclusions:
Forceful expiratory maneuvers with or without a brief (<2s) pause at TLC produce comparable PEF values in test-retest sessions.
Related Concept Videos
Pulmonary Function Tests
Pulmonary Function Tests are crucial diagnostic tools for assessing respiratory function, particularly in patients with chronic respiratory disorders. They comprehensively evaluate lung volumes, ventilatory function, breathing mechanics, diffusion, and gas exchange. These tests help diagnose pulmonary diseases and play a significant role in monitoring disease progression, evaluating disability, and assessing response to therapy.