Related Experiment Video
Updated: Jul 15, 2026

A Model to Simulate Clinically Relevant Hypoxia in Humans
Published on: December 22, 2016
Measurement of carboxyhemoglobin and methemoglobin by pulse oximetry: a human volunteer study
Steven J Barker1, Jeremy Curry, Daniel Redford
1Department of Anesthesiology, University of Arizona College of Medicine, P O Box 245114, Tucson, Arizona 85724-5114, USA. sjbarker@u.arizona.edu
Background:
A new eight-wavelength pulse oximeter is designed to measure methemoglobin and carboxyhemoglobin, in addition to the usual measurements of hemoglobin oxygen saturation and pulse rate. This study examines this device's ability to measure dyshemoglobins in human volunteers in whom controlled levels of methemoglobin and carboxyhemoglobin are induced.
Methods:
Ten volunteers breathed 500 ppm carbon monoxide until their carboxyhemoglobin levels reached 15%, and 10 different volunteers received intravenous sodium nitrite, 300 mg, to induce methemoglobin. All were instrumented with arterial cannulas and six Masimo Rad-57 (Masimo Inc., Irvine, CA) pulse oximeter sensors. Arterial blood was analyzed by three laboratory CO-oximeters, and the resulting carboxyhemoglobin and methemoglobin measurements were compared with the corresponding pulse oximeter readings.
Results:
The Rad-57 measured carboxyhemoglobin with an uncertainty of +/-2% within the range of 0-15%, and it measured methemoglobin with an uncertainty of 0.5% within the range of 0-12%.
Conclusion:
The Masimo Rad-57 is the first commercially available pulse oximeter that can measure methemoglobin and carboxyhemoglobin, and it therefore represents an expansion of our oxygenation monitoring capability.
More Related Videos
04:20Integration of Brain Tissue Saturation Monitoring in Cardiopulmonary Exercise Testing in Patients with Heart Failure
Published on: October 1, 2019
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
Related Concept Videos
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Hypoxia
Types of Hypoxia
There are four primary types of hypoxia, each resulting from a different cause:
1. Anemic hypoxia: This type occurs due to insufficient oxygen delivery caused by a lack of red blood cells (RBCs) or RBCs with abnormal or...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...