Quantitative Analysis of Vacuum Induction Melting by Laser-induced Breakdown Spectroscopy
Published on: June 10, 2019
Spatially resolved, laser-induced breakdown spectroscopy, development, and application for the analysis of Al and Si
Suh-Jen Jane Tsai1, Shi-Yang Chen, Yu-Sheng Chung
1Department of Applied Chemistry, Providence University, Taichung Hsien, Taiwan, R.O.C.
Abstract:
Spatially resolved laser-induced breakdown spectroscopy (SRLIBS) was used for the characterization of aluminum and silicon in nickel-based alloys. The very low invasiveness of the technique was one of the figures of merit of LIBS; however, the relative complexity of the instrument often hindered the widely acceptance of LIBS. Spatially resolved LIBS could provide accuracy and precision comparable to those obtained with temporally resolved LIBS (TRLIBS). In the nongated spatially resolved LIBS, the maximum atomic emission could be obtained with relative low continuum background emission at optimum observation spatial position. The study was done with a Nd:YAG laser at 532 nm, 3.0 mJ laser energy, and 0.2 mbar in argon. The experimental results obtained under optimum conditions were compared to those obtained with TRLIBS. SRLIBS gave reliable results without the tedious optimization of the delay time and gate width.
More Related Videos
09:40Measurement and Analysis of Atomic Hydrogen and Diatomic Molecular AlO, C2, CN, and TiO Spectra Following Laser-induced Optical Breakdown
Published on: February 14, 2014
10:17Laser-induced Breakdown Spectroscopy: A New Approach for Nanoparticle's Mapping and Quantification in Organ Tissue
Published on: June 18, 2014