Fluorescence lifetime correlation spectroscopy combined with lifetime tuning: new perspectives in supported
Ales Benda1, Veronika Fagul'ová, Alexander Deyneka
1J. Heyrovský Institute of Physical Chemistry, Academy of Sciences of the Czech Republic, Dolejskova 3, 18223 Praha 8, Czech Republic.
Abstract:
A new concept based on fluorescence lifetime correlation spectroscopy (FLCS) is presented allowing the simultaneous determination of diffusion coefficients of identical molecules located in different environments. The difference in fluorescence lifetimes, which is the main prerequisite for FLCS, is reached by locating one population of the dye close to a light-absorbing surface. Since such surfaces quench fluorescence, the fluorescence lifetime of chromophores located close to these surfaces can be tuned in a specific manner. This approach has been demonstrated for a BODIPY-tail-labeled lipid in supported phospholipid bilayers (SPBs) as well as in phospholipid multilayers adsorbed onto solid supports. In particular, the effect of the solid support type on the fluorescence lifetime as well as its dependence on the BODIPY-support distance has been characterized and verified by theoretical considerations based on precise determination of refractive indices of the used supports. While the fluorescence lifetime of BODIPY dye is 5.6 ns in small unilamellar vesicles (SUVs) composed of 1,2-dioleoyl-sn-glycero-3-phosphocholine (DOPC) and 1,2-dioleoyl-sn-glycero-3-[phospho-L-serine] (DOPS), the lifetime is 1.8 ns in DOPC/DOPS SPBs adsorbed onto ITO-covered glass or 3.0 ns in a DOPC/DOPS monolayer adsorbed onto seven 1,2-dipalmitoyl-sn-glycero-3-phosphate (DPPA) layers on oxidized silicon. Using these particular systems, we demonstrated that FLCS enables one to characterize simultaneously two-dimensional lipid diffusion in the planar lipid layers and three-dimensional vesicle diffusion in bulk above the lipid layers using single dye labeling. The autocorrelation functions obtained by this new approach do agree with those obtained by standard FCS on isolated SPBs or vesicles. Possible applications of this virtual two-channel measurement using single dye labeling as well as one detection channel are discussed.
Related Concept Videos
Fluorescence and Phosphorescence: Instrumentation
Protein Dynamics in Living Cells
Fluorescent recovery after photobleaching (FRAP) is a fluorescent-protein-based detection technique used to quantify protein movement rates within the cell. This method exposes a small portion of the cell to an intense laser beam. The laser beam causes permanent photobleaching of the fluorophore-tagged proteins in the exposed region. As the bleached...