[The Fabian prosthesis: our experience in 13 cases].

J A Benítez Navío1, M Arrizabalaga Moreno, T Casanueva Luis

  • 1Servicio de Urología del Hospital Ntra. Sra. de Alarcos, Ciudad Real, España.

Archivos Espanoles De Urologia
|November 1, 1990
PubMed
Summary

The spiral intraurethral prosthesis showed poor results and high removal rates in patients with prostatic obstruction. Further research is needed for alternative intraurethral devices.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

[Urogenital alterations in hereditary and sporadic neurodegenerative ataxias].

Neurologia (Barcelona, Spain)·2005
Same author

[Therapeutic options for patients with localized prostatic carcinoma: our experience with 454 patients].

Actas urologicas espanolas·2004
Same author

[Treatment approach for vesicogenital fistula. Retrospective analysis of our data].

Actas urologicas espanolas·2003
Same author

[Prostate cancer in the Community of Madrid in the year 2000. III. Study of tumor extent].

Actas urologicas espanolas·2003
Same author

[Prostate cancer in the Community of Madrid in the year 2000. IV. Treatment].

Actas urologicas espanolas·2003
Same author

[The prostate cancer in the community of Madrid in 2000 I.- Incidence].

Actas urologicas espanolas·2003

Area of Science:

  • Urology
  • Medical Devices

Context:

  • Prostatic obstruction syndrome often requires permanent bladder catheterization.
  • High-risk surgical candidates and those awaiting surgery need alternative solutions.

Purpose:

  • To evaluate the one-year outcomes of the spiral intraurethral prosthesis in patients with prostatic obstruction.
  • To analyze micturition hydrodynamics, continence mechanisms, and identify reasons for poor outcomes.

Summary:

  • A one-year follow-up of 13 patients implanted with a spiral intraurethral prosthesis was conducted.
  • The prosthesis required removal in 6 cases (46%), with calcification being one reason.
  • Poor flowmetry results and incontinence were observed, prompting the abandonment of the metal prosthesis.

Impact:

Related Experiment Videos

  • The spiral intraurethral prosthesis demonstrated suboptimal efficacy and safety in this patient cohort.
  • Findings suggest limitations in current intraurethral device technology for prostatic obstruction.