Wavelet-based tachycardia discrimination in ICDs: Impact of posture and electrogram configuration
Thomas Wolber1, Christian Binggeli, Johannes Holzmeister
1Cardiovascular Center, Cardiology, University Hospital Zurich, Zurich, Switzerland. thomas.wolber@usz.ch
The wavelet algorithm for implantable cardioverter defibrillators (ICDs) shows minor variations in morphology scores across body positions with RV coil-can, RV coil-SVC coil, and tip-ring configurations. However, the SVC coil-can configuration should be avoided due to considerable score variations.
Area of Science:
- Cardiology
- Biomedical Engineering
- Medical Devices
Background:
- Inappropriate therapy delivery is a significant concern in implantable cardioverter defibrillator (ICD) management.
- A novel morphology-based algorithm (wavelet feature) aids in differentiating ventricular and supraventricular tachycardia.
- The study investigates the performance of this wavelet algorithm across various electrogram (EGM) configurations and body positions.
Purpose of the Study:
- To evaluate the performance of the wavelet algorithm in differentiating tachyarrhythmias.
- To assess the impact of different electrogram (EGM) configurations on wavelet algorithm accuracy.
- To determine the influence of body position changes on EGM morphology and algorithm performance.
Main Methods:
- Twenty-eight patients with single-chamber Medtronic ICDs were included.
- Electrogram (EGM) templates were acquired in a supine position using various configurations: RV coil-can, RV coil-SVC coil, tip-ring, and SVC coil-can.
- Morphologic similarity (match percentage) was compared between supine templates and EGMs recorded in different body positions (supine, lateral, sitting, standing, walking).
Main Results:
- A total of 9,775 intracardiac EGMs were analyzed.
- Median match percentages varied by configuration: RV coil-can (88%), RV coil-SVC coil (88%), tip-ring (82%), and SVC coil-can (73%).
- Correct classification rates (>=70% match) were significantly higher for RV coil-can (95%), RV coil-SVC coil (91%), and tip-ring (91%) compared to SVC coil-can (58%).
Conclusions:
- Wavelet-based morphology scores in ICDs are sensitive to body position changes.
- The RV coil-can, RV coil-SVC coil, and tip-ring configurations demonstrate relatively minor variations in morphology scores.
- The SVC coil-can configuration exhibits considerable score variability and should be avoided when utilizing the wavelet algorithm.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and the T...
Dysrhythmias V: Evaluating Dysrhythmias
Dysrhythmias III: Characteristics of Dysrhythmias
Dysrhythmias II: Classification of Tachyarrhythmias
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias