A clinical evaluation of proview pressure phosphene tonometry in children

Ida Chung1, Amelia Bartolone, William H Swanson

  • 1Department of Clinical Sciences, State University of New York, State College of Optometry, New York, New York, USA.

Insights

Proview pressure phosphene tonometry (PPPT) is a noninvasive method for measuring intraocular pressure (IOP) in children. While generally repeatable, PPPT showed significant variability and is not comparable to Goldmann tonometry (GT).

Area of Science:

  • Ophthalmology
  • Optometry
  • Biomedical Engineering

Background:

  • Intraocular pressure (IOP) measurement is crucial for diagnosing and managing glaucoma.
  • Current methods like Goldmann tonometry (GT) require eye drops and direct corneal contact.
  • Noninvasive tonometry methods are sought for improved patient comfort and accessibility, especially in pediatric populations.

Purpose of the Study:

  • To compare intraocular pressure (IOP) measurements obtained in children using Proview pressure phosphene tonometry (PPPT) and Goldmann tonometry (GT).
  • To assess the reliability, validity, and patient preference for PPPT compared to GT in a pediatric cohort.

Main Methods:

  • A study involving 104 children aged 5-12 years at the University Optometric Center/SUNY College of Optometry.
  • Participants were randomized to undergo either PPPT or GT first, with two measurements attempted per eye using each instrument.
  • A subgroup of 41 participants indicated their preference between the two tonometry methods.

Main Results:

  • PPPT had a higher coefficient of repeatability (3-4 mm Hg) compared to GT (1 mm Hg).
  • Mean IOP measurements were 4 mm Hg higher with PPPT than GT, with increasing discrepancies at higher IOPs.
  • 56% of surveyed children preferred PPPT, 24% had no preference, and 20% preferred GT.

Conclusions:

  • PPPT measurements are repeatable within a few mm Hg for most children, but substantial variability exists for some.
  • The mean difference of 4 mm Hg and wide confidence intervals indicate PPPT is not comparable to GT for IOP measurement.
  • Despite comparability issues, PPPT's noninvasive nature and pediatric preference suggest potential as a portable screening tool, warranting further investigation in diverse patient groups.