Related Experiment Video
Updated: Jul 18, 2026

Leveraging Turbidity and Thromboelastography for Complementary Clot Characterization
Published on: June 4, 2020
Diagnosis of early coagulation abnormalities in trauma patients by rotation thrombelastography
L Rugeri1, A Levrat, J S David
1Laboratory of Haemostasis, Edouard Herriot Hospital, Hospices Civils de Lyon and Claude Bernard University, Lyon, France. lucia.rugeri@chu-lyon.fr
Background:
Reagent-supported thromboelastometry with the rotation thrombelastography (e.g. ROTEM) is a whole blood assay that evaluates the visco-elastic properties during blood clot formation and clot lysis. A hemostatic monitor capable of rapid and accurate detection of clinical coagulopathy within the resuscitation room could improve management of bleeding after trauma.
Objectives:
The goals of this study were to establish whether ROTEM correlated with standard coagulation parameters to rapidly detect bleeding disorders and whether it can help to guide transfusion.
Methods:
Ninety trauma patients were included in the study. At admission, standard coagulation assays were performed and ROTEM parameters such as clot formation time (CFT) and clot amplitude (CA) were obtained at 15 min (CA(15)) with two activated tests (INTEM, EXTEM) and at 10 min (CA(10)) with a test analyzing specifically the fibrin component of coagulation (FIBTEM).
Results:
Trauma induced significant modifications of coagulation as assessed by standard assays and ROTEM. A significant correlation was found between prothrombin time (PT) and CA(15)-EXTEM (r = 0.66, P < 0.0001), between activated partial thromboplastin time and CFT-INTEM (r = 0.91, P < 0.0001), between fibrinogen level and CA(10)-FIBTEM (r = 0.85, P < 0.0001), and between platelet count and CA(15)-INTEM (r = 0.57, P < 0.0001). A cutoff value of CA(15)-EXTEM at 32 mm and CA(10)-FIBTEM at 5 mm presented a good sensitivity (87% and 91%) and specificity (100% and 85%) to detect a PT > 1.5 of control value and a fibrinogen less than 1 g L(-1), respectively.
Conclusions:
ROTEM is a point-of-care device that rapidly detects systemic changes of in vivo coagulation in trauma patients, and it might be a helpful device in guiding transfusion.
Related Concept Videos
Venous Thrombosis III: Interprofessional Care
Coagulation
During the coagulation phase, clotting factors, or procoagulants, play a vital role in initiating and progressing the coagulation cascade. This cascade is a series of reactions...
Venous Thrombosis II: Clinical Manifestations and Diagnostic Studies
Extrinsic and Intrinsic Pathways of Hemostasis
The Extrinsic Pathway
The extrinsic pathway of coagulation is typically initiated by tissue damage that exposes blood to tissue factor (TF), a protein released by the damaged tissue cells outside the blood vessels—this interaction with TF triggers biochemical reactions involving specific clotting factors. The key player here is Factor VII, which forms a...
Venous Thrombosis IV: Nursing Management
Clot Retraction and Fibrinolysis