Related Experiment Video
Updated: Jul 18, 2026

Expired CO2 Measurement in Intubated or Spontaneously Breathing Patients from the Emergency Department
Published on: January 29, 2011
Danger of helmet continuous positive airway pressure during failure of fresh gas source supply
Nicolò Patroniti1, Maurizio Saini, Alberto Zanella
1Department of Surgical Sciences and Intensive Care, University of Milan-Bicocca, Ospedale San Gerardo Nuovo dei Tintori, via Donizetti 106, 20052, Monza, Italy. nicolo.patroniti@unimib.it
Objective:
To assess the behavior of different helmets after discontinuation of fresh gas flow by disconnection at the helmet inlet, flow generator, or gas source.
Design And Setting:
Randomized physiological study in a university research laboratory.
Patients:
Five healthy volunteers.
Intervention:
CPAP (FIO2 50%, PEEP 5 cmH2O) delivered in random sequence with three different helmets: 4Vent (Rüsch), PN500 (Harol), CaStar (StarMed) with antisuffocation valve open or locked. For each helmet all three disconnections were randomly employed up to 4 min.
Measurements And Results:
During flow disconnection we measured: respiratory rate and tidal volume by respitrace; inspiratory and expiratory CO2 concentration, and FIO2 from a nostril; SpO2 by pulse oxymetry. Independently of the site of disconnection we observed a fast increase in CO2 rebreathing and minute ventilation, associated with a decrease in inspired O2 concentration. In the absence of an operational safety valve, larger helmet size and lower resistance of the inlet hose resulted in slower increase in CO2 rebreathing. The presence of the safety valve limited the rebreathing of CO2, and the increase in minute ventilation but did not protect from a decrease in FIO2 and loss of PEEP.
Conclusions:
While the use of a safety valve proved effective in limiting CO2 rebreathing, it did not protect from the risk of hypoxia related to decrease in FIO2 and loss of PEEP. In addition to a safety antisuffocation valve, a dedicated monitoring and alarming systems are needed to employ helmet CPAP safely.
Related Concept Videos
Cardiopulmonary Resuscitation II: ACLS Airway Management
Pneumothorax-I
Pneumothorax can be even further classified as spontaneous, traumatic, and tension pneumothorax.
Acute Respiratory Failure-III
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals.
Acute Respiratory Failure-II
The underlying physiological abnormalities that contribute to hypoxemic respiratory failure include:
Acute Respiratory Failure-I
Definition: It is defined by specific criteria based on blood gas measurements. Hypoxemia happens when the partial pressure of oxygen (PaO2) falls below 60 mmHg. At the same time,...