Related Experiment Video
Updated: Jul 18, 2026

Evaluation of Capillary and Other Vessel Contribution to Macular Perfusion Density Measured with Optical Coherence Tomography Angiography
Published on: February 18, 2022
Spatial refractive index measurement of porcine artery using differential phase optical coherence microscopy
Jeehyun Kim1, Digant P Davé, Christopher G Rylander
1Beckman Laser Institute, University of California, Irvine, California, USA. jeehk@uci.edu
Background And Objectives:
We describe a methodology to record spatial variation of refractive index of porcine renal artery using differential phase optical coherence microscopy (DP-OCM).
Study Design/Materials And Methods:
The DP-OCM provides quantitative measurement of thin specimen phase retardation and refractive index by measuring optical path-length changes on the order of a few nanometers and with a lateral resolution of 3 microm. The DP-OCM instrumentation is an all-fiber, dual-channel Michelson interferometer constructed using a polarization maintaining (PM) fiber.
Results:
Two-dimensional en face dual-channel phase images are taken over a 150 x 200 microm region on a microscopic slide, and the images are reconstructed by plotting a two-dimensional refractive index map as the OCM beam is moved across the sample.
Conclusions:
Because the DP-OCM can record transient changes in the optical path-length, the system may be used to record quantitative optical path-length alterations of tissue in response to various stimuli. A fiber-based DP-OCM may have the potential to substantially improve in vivo imaging of individual cells for a variety of clinical diagnostics, and monitoring applications.