[Simulation and experimental study of cone artifact in spiral CT]
Yi-qun Lin1, Ke-cheng Yang, Xiao-e Yu
1School of Biomedical Engineering, Southern Medical University, Guangzhou 510515, China. linyiqun@fimmu.com
Objective:
To study regular patterns of cone artifact resulted from interpolation algorithm of spiral CT.
Methods:
Based on the principle of interpolation algorithm and back-projection reconstruction, a mathematical model of the reconstructed image was established to clarify the relation of the scanning parameters and the characteristics of the scanned object with the cone artifact. Experiments were carried out by a set of acrylic phantoms on siemens plus 4 CT scanners.
Results:
The artifact in the image was directly proportional to the table increment per gantry rotation of the scanner, and was positively correlated to the tangent of half cone-angle and inversely to the radii in the reconstruction plane of the phantom. The theoretical analysis was validated by experimental results.
Conclusion:
The cone artifact is related to the scanning parameters and the characteristics of the scanned object.
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...