Optimizing Mouse Urodynamic Techniques for Improved Accuracy
Published on: June 7, 2024
A comparative study of water perfusion catheters and microtip transducer catheters for urethral pressure measurements
Annette Kuhn1, Charles W Nager, Emma Hawkins
1Department of Urogynaecology, Frauenklinik, Effingerstr.102, Inselspital, 3011, Bern, Switzerland. annette.kuhn@insel.ch
Abstract:
The aim of this study was to compare the maximum urethral closure pressure (MUCP) measures with two different techniques: water perfused catheter and microtip transducer catheters with respect to reproducibility and comparability for urethral pressure measurements. Eighteen women with stress urinary incontinence had repeat static urethral pressure profilometry on a different day using a dual microtip transducer and water perfused catheter (Brown and Wickham). The investigators were blinded to the results of the other. The microtip measurements were taken in the 45 degrees upright sitting position with the patient at rest at a bladder capacity of 250 ml using an 8 Fr Gaeltec double microtip transducer withdrawn at 1 mm/s, and the transducer was orientated in the three o'clock position. Three different measures were taken for each patient. Three water perfusion measurements were performed with the patient at rest in the 45 degrees upright position at a bladder capacity of 250 ml using an 8 Fr BARD dual lumen catheter withdrawn at 1 mm/s. The mean water perfusion MUCP measure was 26.1 cm H2O, significantly lower than the mean microtip measure of 35.7 cm H2O. The correlation coefficient comparing each water perfusion measurement with the other water perfusion measures in the same patient was excellent, at 0.95 (p = 0.01). Correlation coefficient comparing each microtip measure with the other microtip measure in the same patient was also good, ranging from 0.70 to 0.80. This study confirms that both water perfusion catheters and microtip transducers have excellent or very good reproducibility with an acceptable intraindividual variation for both methods.
More Related Videos
Related Concept Videos
Urodynamic Studies: Uroflowmetry
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Imaging Studies V: Intravenous Urography and Retrograde Pyelography