Comparison of Bayer Advia Centaur immunoassay results obtained on samples collected in four different Becton
A Morovat1, T S James, S D Cox
1Department of Clinical Biochemistry, John Radcliffe Hospital, Oxford OX3 9DU, UK. reza.moravat@orh.nhs.uk
Background:
Medicines and Healthcare products Regulatory Agency's (MHRA's) Medical Device Alert MDA/2004/048 described bias in some endocrine test results obtained on a few immunoassay platforms, particularly the Bayer Advia Centaur instrument, when using blood specimens collected into Becton Dickinson (BD) Vacutainer SSTII Advance tubes. As users of BD tubes and the Advia Centaur instrument, we addressed our concerns about the quality of the results that we had previously reported by undertaking an independent study.
Method:
We compared the results of 15 immunoassays performed on Bayer Advia Centaur using blood specimens collected into four different BD Vacutainer tubes (plain, old and newly released BD SSTII Advance, and BD PSTII).
Results:
Compared with plain tubes, old SSTII Advance tube results showed no bias for testosterone, CA15-3, follicle-stimulating hormone and folate assays, but gave a positive bias for cortisol and a negative bias for vitamin-B12. Compared with plain tubes, BD PSTII tubes gave no significant bias for thyroid function tests, prolactin, parathyroid hormone, and CA125, but gave a negative bias for steroid assays, and a positive bias for gonadotrophins. The results obtained using new BD SSTII Advance tubes were generally comparable with those on plain tubes.
Conclusions:
Only for cortisol did our findings support the bias described by MHRA. Based on our results, apart from vitamin-B12 and possibly cortisol, there may have been no significant influence on clinical decisions as a result of using the old BD SSTII Advance specimen tubes. New BD SSTII Advance tubes and plain tubes give generally comparable results. BD PSTII tubes should not be used for steroid hormone measurements on the Bayer Advia Centaur instrument.