Related Experiment Video
Updated: Jul 17, 2026

Dried Blood Spots - Preparing and Processing for Use in Immunoassays and in Molecular Techniques
Published on: March 13, 2015
Effects of storage at -20 degrees C on ischaemia-modified albumin results
1Department of Clinical Biochemistry, North Bristol NHS Trust, Frenchay Hospital, Bristol BS16 1LE, UK. robert.beetham@nbt.nhs.uk
Background:
Ischaemia-modified albumin (IMA) is being studied as a new marker for reversible ischaemia in patients presenting with possible cardiac chest pain. The conditions under which samples are stored prior to analysis may be critical in influencing the analytical result and hence the cut-off used in any particular study.
Methods:
Sixty-eight samples taken during a study assessing the performance of IMA for risk stratification in patients presenting with possible cardiac chest pain were analysed both within 2.5 h of collection and after periods of storage at -20 degrees C.
Results:
Samples stored at -20 degrees C yielded IMA values on average 3 units higher than those analysed within 2.5 h (mean 90.5 vs. 87.5; P < 0.00001). A Bland-Altman plot showed that the difference was not concentration dependent.
Conclusions:
These results indicate that decision cut-offs will be influenced by conditions of sample storage prior to IMA analysis, and that these should be stated in detail for each study.

