Air microbubbles have been investigated recently at high magnetic field strength (2 Tesla or greater) as potential MR susceptibility contrast agents. We used a phantom to measure their susceptibility at 1.5 T to clarify their usefulness for this purpose. The phantom, filled with fresh Levovist suspension at 4 different doses (67 to 125 mg/mL), was continuously scanned with a gradient-echo technique at a temporal resolution of 10 s. The transverse relaxation increase (R2*) by microbubbles demonstrated a time course of exponential decay at each dose (time-constant, 39 to 57 s). The dependency of R2* on microbubble volume fraction was linear, with a slope of 89 s-1 per percentage microbubble volume fraction. Our study represents the first step towards applying microbubbles as susceptibility contrast agents at 1.5 T.