[Basic evaluation of the new pulse sequence for simultaneous acquisition of T1- and T2-weighted images]
Atsunari Kurose1, Tohru Takahashi, Tae Ohishi
1Department of Radiology, Nippon Medical School Musashikosugi Hospital.
Abstract:
A novel pulse sequence that enables simultaneous acquisition of T1-weighted (T1W) and T2-weighted (T2W) images is presented. In this new technique, the inversion recovery (IR) pulse of conventional fast inversion recovery (Fast IR) is replaced with a pulse train that consists of a fast spin echo (FSE) and 180 (y) +90 (x) for driven inversion (DI). By using a shorter TI and independent k-space ordering, the first part of the sequence provides T2W images and the second part provides T1W images, thereby enabling simultaneous acquisition in a single scan time comparable to that of Fast IR. Signal simulation also was conducted, and this was compared with conventional scanning techniques using normal volunteers. In the human studies, both T1W and T2W images showed the same image quality as conventional images, suggesting the potential for this technique to replace the combination of Fast IR and T2W FSE for scan-time reduction.