Related Experiment Video
Updated: Jul 18, 2026

Simultaneous Scalp Electroencephalography (EEG), Electromyography (EMG), and Whole-body Segmental Inertial Recording for Multi-modal Neural Decoding
Published on: July 26, 2013
Removal of eye movement artefacts from single channel recordings of retinal evoked potentials using synchronous
A C Fisher1, W El-Deredy, R P Hagan
1Department of Clinical Engineering, Royal Liverpool University Hospital, Medical School, Duncan Building, Prescot Street, Liverpool L78XP, UK. a.c.fisher@liv.ac.uk
Abstract:
A system is described for the removal of eye movement and blink artefacts from single channel pattern reversal electroretinogram recordings of very poor signal-to-noise ratios. Artefacts are detected and removed by using a blind source separation technique based on the jadeR independent component analysis algorithm. The single channel data are arranged as a series of overlapping time-delayed vectors forming a dynamical embedding matrix. The structure of this matrix is constrained to the phase of the stimulation epoch: the term synchronous dynamical embedding is coined. A novel method using a marker channel with a non-independent synchronous feature is employed to identify the single most relevant source estimation for reconstruction and signal recovery. This method is non-lossy, all underlying signal being recovered. In synthetic datasets of defined noise content and in standardised real data recordings, the performance of this technique is compared to conventional fixed-threshold hard-limit rejection. The most significant relative improvements are achieved when movement and blink artefacts are greatest: no improvement is demonstrable for the random noise only situation.