Related Experiment Video
Updated: Jul 18, 2026

The Use of a β-lactamase-based Conductimetric Biosensor Assay to Detect Biomolecular Interactions
Published on: February 1, 2018
Organophosphorus hydrolase multilayer modified microcantilevers for organophosphorus detection
Chandana Karnati1, Hongwei Du, Hai-Feng Ji
1Department of Chemistry, Institute for Micromanufacturing, Louisiana Tech University, Ruston, LA 71272, United States.
Abstract:
We report a biosensor based on organophosphorus hydrolase (OPH) multilayer modified microcantilever (MCL) for detection of organophosphorus compounds (OPs). The assay is based on substrate-dependent bending of the OPH functionalized MCLs. The cantilever bending amplitude at equilibrium was a function of the concentration of paraoxon with the dynamic range extending from 10(-7) to 10(-3)M. The lower detection limit of approximately 10(-7)M for paraoxon was an order of magnitude better than the OPH-based potentiometric and optical biosensors based on pH modulation. There was a good intra-sensor and an acceptable inter-sensor reproducibility as evidenced by the standard errors of 5% and 15%, respectively. OPs measured using this technique included parathion and diisopropyl fluorophosphate (DFP) in the order of paraoxon>DFP>parathion. The conformational change of the OPH was most likely the main origin of MCL bending.
Related Concept Videos
Microbial Biosensors
Anticholinesterase Agents: Poisoning and Treatment
Irreversible agents form a strong bond with the cholinesterase enzyme, making it inactive. The breakdown of the phosphorylated enzyme is slower than the...
High-Performance Liquid Chromatography: Types of Detectors