Related Experiment Video
Updated: Jul 18, 2026

Laser-free Hydroxyl Radical Protein Footprinting to Perform Higher Order Structural Analysis of Proteins
Published on: June 4, 2021
Development of a liquid chromatography/electrospray ionization tandem mass spectrometric method for the determination
Fen Yang1, Ruiping Zhang, Jiuming He
1Institute of Materia Medica, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing 100050, China.
Abstract:
A rapid method combining liquid chromatography with electrospray ionization tandem mass spectrometry (LC/ESI-MS/MS) was developed for the determination of the hydroxyl radical (.OH). .OH generated via Fenton reaction was spin-trapped by 5,5-dimethyl-1-pyrroline N-oxide (DMPO) and then analyzed by LC/ESI-MS/MS in multiple reaction monitoring (MRM) mode, using N-methyl-2-pyrrolidone (NMP) as the internal standard. The peak area ratio of DMPO-OH to NMP positively correlated with the concentration of .OH. The relative standard deviation (RSD) of the method was 1.13% (n = 8). The present method was successfully applied to evaluate the .OH scavenging capacity of several phenolic acids.
Related Concept Videos
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
Mass Spectrometry: Molecular Fragmentation Overview
One type of fragmentation pattern is the cleavage of a single bond in the molecular ion. The cleavage leads to a radical and a cation. The cleavage can occur at...
Radical Reactivity: Overview