Related Experiment Video
Updated: Jul 18, 2026

Dynamic Lung Tumor Tracking for Stereotactic Ablative Body Radiation Therapy
Published on: June 7, 2015
Four-dimensional measurement of lung tumor displacement using 256-multi-slice CT-scanner
Shinichiro Mori1, Masahiro Endo, Shuhei Komatsu
1Research Center for Charged Particle Therapy, National Institute of Radiological Sciences, Chiba, Japan. smori1@partners.org
Abstract:
The concept of internal target volume is of marked importance for radiotherapy to lung tumors as respiration-induced motion is important. Individualized assessment of motion is required as tumor site may not predict the extent or pattern of tumor motion. We performed volumetric cine scanning using the 256-multi-slice CT (256MSCT) to study tumor motion during free breathing in 14 inpatients who were treated with carbon-ion radiotherapy. Motion assessment in 16 respiratory phases of the cine CT revealed most tumors to show hysteresis-like behavior. Isocenter displacement between peak exhalation and inhalation for the average of the right and left lungs were 7 mm, 7 mm and 15 mm for the upper, middle and lower lobes, respectively. Cine CT with the 256MSCT improved the evaluation of tumor displacement and overcomes some of the limitations associated with current CT methods. Volumetric cine CT data provides useful data on motion for planning in all radiation approaches for lung tumors.

