Collection of nonpolar organic compounds from ambient air using polyurethane foam-granular adsorbent sandwich
M T Zaranski1, G W Patton, L L McConnell
1Department of Chemistry, University of South Carolina, Columbia 29208.
Abstract:
Glass cartridges containing 6-10 g of Tenax-GC or 15 g of XAD-2 resin packed between two slices of polyurethane foam (PUF) were used in a General Metal Works PS-1 high-volume sampler to collect chlorobenzenes (CBs) containing three to six chlorine atoms, hexachlorocyclohexanes (HCHs), and two-ring aromatic hydrocarbons from 35-385 m3 of air. Laboratory experiments were run by vaporizing known quantities of analytes into a clean airstream for sampling by the PS-1 system at 20 degrees C. Collection efficiencies, determined from mass balance of the quantities introduced and recovered, ranged from 70 to 120% for individual compounds and averaged 93% overall. Penetration of analytes to backup adsorbent traps showed an inverse correlation to vapor pressure. The method was used to collect the above compounds from urban and rural air.
Related Concept Videos
Sampling Methods: Sample Types
Solid samples include a variety of substances, such as sediments from water bodies, soil, metals, and biological tissues. Two standard methods for extracting sediments from water bodies are grab sampling and piston coring. Grab sampling involves using a device to collect a discrete sediment sample from the bottom of a water body with minimal disturbance. Grab samples do not always represent the entire area due to...