[Factors related to the differences in blood pressure values assessed by auscultatory or oscillometric methods]
María Rosa Oltra1, Antonio Vicente, José Vicente Lozano
1Unidad de Hipertensión Arterial, Servicio de Medicina Interna, Hospital Clínic Universitari, Universitat de València, Valencia, Spain.
Background And Objective:
Measurement of blood pressure by using the auscultatory method coupled with mercury sphygmomanometer is rapidly being replaced for monitor using the oscillometric one. Discrepancies between the two methods and the factors related to them were analyzed in a large subset of adults.
Material And Method:
Blood pressure values were obtained sequentially in the same arm by using a mercury sphygmomanometer and a Spacelabs monitor devices. The relationship between both kinds of measurements was assessed by the Pearson's correlation coefficients and the concordance by using the Bland and Altman method. Differences between the two methods were calculated and the factors related to the differences were sought by multiple regression models.
Results:
A total of 1,742 adult subjects were analyzed. Pearson's correlation coefficients were 0.927, 0.922, 0.886, 0.962 for SBP, DBP, MBP and PR, respectively. Oscillometric method overestimate and infraestimate SBP values in the two extreme of SBP values, the lowest SBP the highest the values and vice versa. A similar trend was observed for DBP although the differences were minor. Pulse wave amplitude explains the 24% of the variance observed for SBP and arm circumference the 8% for DBP.
Conclusions:
Blood pressure values measured by oscillometric methods are influenced for pulse wave amplitude and by the arm circumference. These need to be taken in account when blood pressure assessment is obtained by using this method.
Related Concept Videos
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Measurement of Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Factors affecting Blood pressure
Physiological Factors: