Abstract:
Twenty-nine patients undergoing left thoracotomy consecutively were intubated with right-sided double-lumen endotracheal tubes in the conventional "blind" fashion. The first 20 received Leyland red-rubber double-lumen tubes, and the subsequent nine received Mallinckrodt polyvinylchloride double-lumen tubes. The patency of the right upper lobe bronchial orifice was then assessed by flexible fiberoptic bronchoscopy. Right upper lobe obstruction occurred in 89% of the polyvinylchloride tube intubations and in 10% of the red-rubber tube intubations.
Related Concept Videos
The Bronchial Tree
The trachea, commonly known as the windpipe, is a tube that connects the larynx (voice box) to the bronchi. At a point called the carina, it bifurcates into two primary bronchi. The right primary bronchus is wider, shorter, and more vertical than the left primary...
Anatomy of Respiratory System II: Lower Respiratory Tract
The Larynx
It is located between the pharynx and the trachea, acts as a passageway for air, and hosts several critical structures, such as the epiglottis, vocal cords, and glottis. The epiglottis acts as a gateway, guiding food to the...
Gross Anatomy of the Lungs
Endoscopic Studies I: Bronchoscopy and Thoracoscopy
Bronchoscopy
Description
Bronchoscopy is a procedure that involves direct visualization of the larynx, trachea, and bronchi for diagnostic and therapeutic purposes. A flexible fiber optic or rigid bronchoscope is used to carry out the procedure. The fiber-optic bronchoscope is more frequently used due to...
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Chronic Obstructive Pulmonary Disease III: Chronic Bronchitis Features