Related Experiment Video
Updated: Jul 18, 2026

Construction and Application of Cerebral Functional Region-Based Cerebral Blood Flow Atlas Using Magnetic Resonance Imaging-Arterial Spin Labeling
Published on: May 31, 2024
Nonlinear magnetic field gradients can reduce SAR in flow-driven arterial spin labeling measurements
Kenneth I Marro1, Donghoon Lee, Outi M Hyyti
1Department of Radiology, Box 357115, University of Washington, Seattle, WA 98195-7115, USA. marro@u.washington.edu
Abstract:
This work describes how custom-built gradient coils, designed to generate magnetic fields with amplitudes that vary nonlinearly with position, can be used to reduce the potential for unsafe tissue heating during flow-driven arterial spin labeling processes. A model was developed to allow detailed analysis of the adiabatic excitation process used for flow-driven arterial water stimulation with elimination of tissue signal (FAWSETS) an arterial spin labeling method developed specifically for use in skeletal muscle. The model predicted that, by adjusting the amplitude of the gradient field, the specific absorption rate could be reduced by more than a factor of 6 while still achieving effective labeling. Flow phantom measurements and in vivo measurements from exercising rat hind limb confirmed the accuracy of the model's predictions. The modeling tools were also applied to the more widely used continuous arterial spin labeling (CASL) method and predicted that specially shaped gradients could allow similar reductions in SAR.