Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018
Safe magnetic resonance scanning of patients with metallic middle ear implants
1ENT Department, Queens Medical Centre, Nottingham, UK. dcwilde@rcsed.ac.uk
Objective:
To determine the MR scanning risk to patients with otologic implants.
Design:
We used a repeated-measures study with an additional control measure to assess two aspects of risk; (i) movement of the device in the magnetic field, and (ii) absorption of energy leading to local heating. We used an ex vivo test method that met with international standards. We measured the effects in a Philips Intera Achieva 3 Tesla (T) MR scanner using a Sense Head 8 channel RF coil.
Setting:
University-based magnetic resonance research facility.
Main Outcome Measures:
Heating or displacement of the stapedectomy pistons.
Results:
No evidence of displacement or heating was found.
Conclusion:
Complying with the ex vivo standard testing protocols, the Schuknecht and McGee wire pistons (device product numbers 140106 and 140108, respectively) were found to be safe in a 3 T MR scanner. These conclusions can be extrapolated to the in vivo case.
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies IV: Magnetic Resonance Imaging