Multicentered European study on safety and effectiveness of the On-X prosthetic heart valve: intermediate follow-up
George M Palatianos1, Axel M Laczkovics, Paul Simon
1Onassis Cardiac Surgery Center, Athens, Greece. palatianos@otenet.gr
Background:
This study was performed to determine the safety and effectiveness of the On-X valve, a novel mechanical valve substitute.
Methods:
Eleven centers participated in a European, multicentered, longitudinal, nonrandomized study of the On-X valve performance. Isolated aortic or mitral valve replacement with an On-X valve was studied in 301 patients. Aortic valve replacement was performed in 184 patients (average follow-up, 5.0 years), whereas mitral valve replacement was performed in 117 patients (average follow-up, 4.4 years).
Results:
In patients with aortic valve replacement, mean transvalvular pressure gradients ranged from 8.3 to 4.7 mm Hg and effective orifice areas from 1.5 to 2.7 cm2, for 19-mm through 25-mm valves, respectively. After mitral valve replacement, mean gradient was 4.2 mm Hg and effective orifice area by pressure half-time was 2.6 cm2 regardless of valve size. Hemolysis was low, with postoperative serum lactate dehydrogenase at 225 +/- 41 IU (mean +/- standard deviation) or 253 +/- 65 IU, after aortic valve replacement or mitral valve replacement, respectively (upper normal value, 250 IU). At 1 year or greater postoperatively, 91.6% of patients after aortic valve replacement and 84.6% after mitral valve replacement were in New York Heart Association functional class I or II. Adverse event rates in percent per patient-year after aortic valve replacement or mitral valve replacement were thromboembolism, 0.88 or 1.76; thrombosis, 0.11 or 0.20; bleeding, 0.77 or 1.96, respectively. Late mortality was 1.97% or 2.55%, respectively.
Conclusions:
At the intermediate follow-up, the On-X valve exhibited improved hemodynamics, low hemolysis with in-range lactate dehydrogenase, and low adverse event rates, particularly in the aortic position.
Related Concept Videos
Mitral Valve Prolapse II: Assessment and Management
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Mitral Stenosis III: Medical Management
Endocarditis III: Medical Management
Acute Coronary Syndrome III: Diagnostic Studies
Mitral Stenosis I: Introduction