Error analysis of ultrasonic tissue doppler velocity estimation techniques for quantification of velocity and strain
Michael J Bennett1, Steve McLaughlin, Tom Anderson
1School of Clinical Sciences and Community Health, Medical and Radiological Sciences, Medical Physics, The Chancellor's Building, 49 Little France Crescent, Edinburgh, EH16 4SB UK. mjb@ee.ed.ac.uk
Abstract:
Recent work in the field of Doppler tissue imaging has focused mainly on the quantification of results involving the use of techniques of strain and strain-rate imaging. These results are based on measuring a velocity gradient between two points, a known distance apart, in the region-of-interest. Although many recent publications have demonstrated the potential of this technique in clinical terms, the method still suffers from low repeatability. The work presented here demonstrates, through the use of a rotating phantom arrangement and a custom developed single element ultrasound system, that this is a consequence of the fundamental accuracy of the technique used to estimate the original velocities. Results are presented comparing the performance of the conventional Kasai autocorrelation velocity estimator with those obtained using time domain cross-correlation and the complex cross-correlation model based estimator. The results demonstrate that the complex cross-correlation model based technique is able to offer lower standard deviations of the velocity gradient estimations compared with the Kasai algorithm.
More Related Videos
12:12Echocardiographic Approaches and Protocols for Comprehensive Phenotypic Characterization of Valvular Heart Disease in Mice
Published on: February 14, 2017
11:50High-frequency High-resolution Echocardiography: First Evidence on Non-invasive Repeated Measure of Myocardial Strain, Contractility, and Mitral Regurgitation in the Ischemia-reperfused Murine Heart
Published on: July 9, 2010
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Ultrasonography
During an ultrasonography procedure, a handheld device called a...
Measurements of Strain