Objective:
To develop a specific, sensitive, reproducible SPE-HPLC method for the determination of 37 drugs in whole blood.
Methods:
With the doxapram as internal standard, Oasis column was used to extract drugs from whole blood. Two kinds of mobile phases were used in this study. Separations were achieved by a LiChrospher 100 RP-C18 (250 mm x 4.0 mm x 5 microm) column kept at 50 degrees C, the DAD detector was set at 230 nm and 250 nm.
Results:
The limit of detection were 1-30 ng/mL. The method showed excellent linearity and the linear correlation coefficient was > or =0.997 98. The relative standard deviation for between-day and within-day assay were <10%.
Conclusion:
The method is effective, simple, reliable and has been used in real cases.
More Related Videos
13:35A Convenient Method for Extraction and Analysis with High-Pressure Liquid Chromatography of Catecholamine Neurotransmitters and Their Metabolites
Published on: March 1, 2018
11:17A Simple Fractionated Extraction Method for the Comprehensive Analysis of Metabolites, Lipids, and Proteins from a Single Sample
Published on: June 1, 2017
Related Concept Videos
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Instrumentation
High-Performance Liquid Chromatography: Elution Process
High-Performance Liquid Chromatography: Types of Detectors
Hepatic Drug Clearance: Effect of Protein Binding
For low-extraction-ratio drugs that are less than 80% protein-bound, minor changes in protein binding...
Extracorporeal Removal of Drugs: Hemoperfusion and Hemofiltration