Inappropriate shock: a failure of SVT discriminators in a dual chamber ICD?
Amin Al-Ahmad1, Angela Tsiperfal, Henry H Hsia
1Cardiac Arrhythmia Service, Stanford University School of Medicine, Stanford, California 94305, USA. aalahmad@cvmed.stanford.edu
Abstract:
Inappropriate shock remains a major issue in patients with implantable cardioverter defibrillators. We examine an inappropriate shock delivered in a patient with atrial tachycardia that conducted 1:1. We reconstruct the device algorithms that led to therapy delivery and discuss programming changes that could be considered.