Calculation of 10 MV x-ray spectra emitted by a medical linear accelerator using the BFGS quasi-Newton method
T Shimozato1, K Tabushi, S Kitoh
1Hamamatsu University School of Medicine, University Hospital, 1-20-1 Handayama, Hamamatsu, Shizuoka 431-3192, Japan. shimo-p@umin.ac.jp
Abstract:
To calculate photon spectra for a 10 MV x-ray beam emitted by a medical linear accelerator, we performed numerical analysis using the aluminium transmission data obtained along the central axis of the beam under the narrow beam condition corresponding to a 3x3 cm2 field at a 100 cm distance from the source. We used the BFGS quasi-Newton method based on a general nonlinear optimization technique for the numerical analysis. The attenuation coefficients, aluminium thicknesses and measured transmission data are necessary inputs for the numerical analysis. The calculated x-ray spectrum shape was smooth in the lower to higher energy regions without any angular components. The x-ray spectrum acquired by the employed method was evaluated by comparing the measurements along the central axis percentage depth dose in a water phantom and by a Monte Carlo simulation code, the electron gamma shower code. The values of the calculated percentage depth doses for a 10x10 cm2 field at a 100 cm source-to-surface distance in a water phantom were obtained using the same geometry settings as those of the water phantom measurement. The differences in the measured and calculated values were less than +/-1.0% for a broad region from the shallow part near the surface to deep parts of up to 25 cm in the water phantom.
Related Concept Videos
Biological Effects of Radiation
Calculation of Electric Flux
Gauss's Law: Problem-Solving