Related Experiment Video
Updated: Jul 17, 2026

Hollow Microneedle-based Sensor for Multiplexed Transdermal Electrochemical Sensing
Published on: June 1, 2012
Application of lactate amperometric sol-gel biosensor to sequential injection determination of L-lactate
Sofia Piedade Gomes1, Martina Odlozilíková, M Gabriela Almeida
1REQUIMTE, Departamento de Química-Física, Faculdade de Farmácia, Universidade do Porto, Rua Anibal Cunha, 164, 4099-030 Porto, Portugal.
Abstract:
This work describes the construction and evaluation of lactate sol-gel biosensors to accomplish the determination of lactate in pharmaceutical products. Lactate oxidase was incorporated in a porous sol-gel film placed onto a platinum-based electrode. Acid and basic catalysis were assessed. When coupled to a sequential injection system (SIA) the biosensor, based on (3-aminopropyl)trimethoxysilane, 2-(3,4-epoxycyclohexyl)ethyl-trimethoxysilane, deionised water, polyethylene glycol 6000 and acid catalyst, presented a range of linearity of 5x10(-5) to 5x10(-3)M. The analytical usefulness of the developed biosensor was evaluated through analysis of commercial pharmaceutical products containing lactate with a sampling rate of 40 samples h(-1). The enzyme remained active for at least 30 days, enabling about 700 determinations without sensitivity decrease.
More Related Videos
13:42Dry Film Photoresist-based Electrochemical Microfluidic Biosensor Platform: Device Fabrication, On-chip Assay Preparation, and System Operation
Published on: September 19, 2017
08:06The Use of a β-lactamase-based Conductimetric Biosensor Assay to Detect Biomolecular Interactions
Published on: February 1, 2018