Analysis of effective radiating area, power, intensity, and field characteristics of ultrasound transducers
Lennart D Johns1, Stephen J Straub, Samuel M Howard
1Department of Physical Therapy, Athletic Training/Sports Medicine Program, Quinnipiac University, Hamden, CT, USA. Lenn.Johns@quinnipiac.edu
Objective:
To characterize the ultrasound fields produced by a cohort of transducers from a single manufacturer via hydrophone and Schlieren technology.
Design:
Descriptive study.
Setting:
Measurement laboratory.
Participants:
Seven same-model ultrasound transducers from a single manufacturer.
Interventions:
Not applicable.
Main Outcome Measures:
Effective radiating area (ERA), total power, spatial average intensity (SAI), beam nonuniformity ratio (BNR), and Schlieren beam widths at 1.0 and 3.3 MHz.
Results:
Values for ERA (1.0 MHz range, 3.62-4.38 cm(2); 3.3 MHz range, 3.74-4.76 cm(2)), total power (1.0 MHz range, 5.0-5.6 W; 3.3 MHz range, 4.7-5.7 W), SAI (1.0 MHz range, 1.2-1.4 W/cm(2); 3.3 MHz range, 1.0-1.5 W/cm(2)), and BNR (1.0 MHz range, 2.79-5.85; 3.3 MHz range, 2.51-4.56) fell within manufacturer's specifications and U.S. Food and Drug Administration (FDA) regulations. Schlieren analysis showed significantly larger beam widths at 3.3 MHz compared with 1.0 MHz and a large degree of variability in the ultrasound fields generated by the different transducers. There were no significant correlations between beam widths and ERA values.
Conclusions:
ERA and total power values in a test cohort exist within a range that met FDA regulations. Individual variability in ERA and total power resulted in 50% variability in SAI. This variability may help explain previous reports of heating differences between transducers.
More Related Videos
09:05Generation and Quantitative Analysis of Pulsed Low Frequency Ultrasound to Determine the Sonic Sensitivity of Untreated and Treated Neoplastic Cells
Published on: July 22, 2015
08:08Evaluating Targeting Accuracy in the Focal Plane for an Ultrasound-guided High-intensity Focused Ultrasound Phased-array System
Published on: March 6, 2019
Related Concept Videos
Ultrasonography
During an ultrasonography procedure, a handheld device called a...
Sound Intensity
Intensity Of Electromagnetic Waves
Energy and Power of a Wave
Waves can also be concentrated or spread out, as characterized by the intensity of the wave. Intensity is directly...
Ultrasound I: Abdominal Ultrasonography
Abdominal ultrasonography, commonly known as abdominal ultrasound, is a vital, non-invasive medical imaging technique widely used in healthcare.
Procedure:
This diagnostic tool allows the clinician to visually inspect internal structures within the abdomen, including vital organs such as the liver, gallbladder, pancreas, kidneys, and spleen.
The abdominal ultrasound process begins with applying a special gel to the patient's skin over the abdomen. This gel enhances the...
Intensity and Pressure of Sound Waves
Unlike the time average of a sinusoidal term, which is zero since it is positive and...