Purpose:
We describe a novel method to oxygenate intraocular irrigation solutions involving an in-line oxygenator.
Methods:
Either lactated Ringer (LR) solution or balanced salt solution (BSS) was oxygenated with the FE390 Stainless In-line Oxygenation Assembly (Beer, Beer, and More Beer, Concord, CA). After running a 100-mL of solution through the in-line oxygenator, oxygen saturation was measured with a dissolved oxygen meter. A control experiment involving the oxygenator without oxygen perfusion was performed. Paired t-tests were used to compare oxygen saturation changes before and after oxygenation.
Results:
In comparison with the original BSS, there was an increase in oxygen saturation of 162 +/- 47% (n = 40; P < 0.05). Without oxygenation perfusion, there was only a 24 +/- 14% (n = 40) increase in oxygen levels in BSS, which was significantly lower than that in treated BSS (P < 0.05). In comparison with the original LR solution, there was an increase in oxygen saturation of 208 +/- 21% (n = 40; P < 0.05). Without oxygenation perfusion, there was only a 21 +/- 9% (n = 40) gain in oxygen saturation in the control LR solution, which was statistically lower as well (P < 0.05).
Conclusion:
The in-line oxygenator is an efficient tool for oxygenating BSS and LR solution. It represents a potential efficient and convenient method to oxygenate irrigating solutions for vitreoretinal surgeries.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Bioreactor Controls-II
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils, connected...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes: