Abstract:
A sensitive solid-phase extraction/high-performance liquid chromatographic method with ultraviolet detection was established for the analysis of salvianolic acid B in rabbit plasma. The analyte was separated on a reversed-phase column with trifluoroacetic acid-methanol-acetonitrile (70:10:20, v/v/v) as mobile phase at a flow rate of 1 mL/min, and ultraviolet detection at 315 nm. The calibration curve for salvianolic acid B was linear over the range 35-1400 microg/L with coefficients of correlation >0.999. The inter-day and intra-day precisions of analysis were <15%, and assay accuracy ranged from 95.3 to 109.1%. This method is suitable for determining salvianolic acid B in plasma and thus investigating the pharmacokinetics of salvianolic acid B.
Related Concept Videos
Methods for Studying Drug Absorption: In situ
The Doluisio method involves perfusing a prepared segment of a rat's small intestine with a solution of radiolabeled drug and a non-absorbable marker. This helps to differentiate between absorbed and non-absorbed drug concentrations. The intestinal segment is connected at both ends using tubing and syringes,...