Intraluminal pressure response in Baerveldt tube shunts: a comparison of modification techniques
1Wake Forest University Eye Center, Wake Forest University School of Medicine, Winston-Salem, NC 27157-1033, USA.
Purpose:
To design an apparatus for in vitro quantification of effects of modifications to nonvalved tube shunts, and to compare the effects of these modifications done to prevent early postoperative hypotony.
Materials And Methods:
A testing apparatus was designed and constructed to measure intraluminal pressure (ILP) during constant infusion of balanced salt solution at 2 microL/min through Baerveldt tube shunts mounted on cadaver eyes. Three different modifications were performed and 3 shunts were used for each modification. The modifications were partial tube occlusion with a 3-0 Supramid suture, tube perforation with a 27-gauge needle on an occluded tube, and a 1.2 mm longitudinal venting slit on an occluded tube.
Results:
The final steady state ILP for the 3-0 Supramid partially occluded tube ranged between 1.6 and 2.8 mm Hg over 3 trials with an average final steady state ILP of 2.0 mm Hg. The 27-gauge needle perforation produced ILPs of 4.6 to 8.2 mm Hg over 3 trials with an average final steady state pressure of 5.9 mm Hg. The 1.2 mm longitudinal venting slit produced an ILP range of 12.6 to 17.3 mm Hg over 3 trials with an average final steady state ILP of 14.8 mm Hg. The differences in final steady state ILP between the 3 modification techniques were statistically significant (P=0.00036).
Conclusions:
The 3-0 Supramid occlusion and the 27-gauge perforation techniques produced hypotony, whereas the 1.2 mm longitudinal venting slit produced acceptable pressures in vitro.
Related Concept Videos
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs during...