Related Experiment Video
Updated: Jul 17, 2026

Optimized Griess Reaction for UV-Vis and Naked-eye Determination of Anti-malarial Primaquine
Published on: October 11, 2019
Very small injected samples to study chloroquine and quinine in human serum using capillary-LC and native
H Ibrahim1, J Bouajila, N Siri
1UMR 152-IRD-Université Paul Sabatier, Toulouse 3, 31062 Toulouse Cedex 9, France.
Abstract:
A comparison between HPLC with conventional fluorescence detection and capillary-LC (microHPLC) with native laser-induced fluorescence (LIF) detection was done to determine chloroquine (CQ) and quinine (Q) in human serum. HPLC experiments were run with parameters of the conventional fluorimeter set at the highest level of sensitivity. Results were compared with those obtained on microHPLC coupled to a ZETALIF (He-Cd 325 nm) detector which provided a 50-fold increase in sensitivity. In microHPLC-LIF injection volumes were 200 nL instead of 10 microL in conventional HPLC. The separation was completed within 3 min (6 min on HPLC). The limit of detection on microHPLC-LIF was 1.9 and 1.3 fmol for CQ and Q, respectively. Both experiments were validated on serum samples. The mean recovery was more than 95% for CQ and Q. The intra- and inter-day precision and accuracy were found to be within the acceptable limits (<10%).