Haemodynamic parameters obtained by transthoracic echocardiography and Swan-Ganz catheter: a comparative study in liver transplant patients
Related Experiment Video
Updated: Jul 17, 2026

Functional Assessment of the Donor Heart During Ex Situ Perfusion: Insights from Pressure-Volume Loops and Surface Echocardiography
Published on: October 11, 2022
Paulo Marcelino1, Nuno Germano, Susan Marum
1Unidade de Cuidados Intensivos, Hospital de Curry Cabral, Centro de Cardiologia da Universidade de Lisboa, Lisboa, Portugal.
Objective:
To compare the haemodynamic measurements of cardiac output (CO), central venous pressure (CVP), pulmonary capillary wedge pressure (Pw) and pulmonary artery systolic pressure (PASP), obtained by Swan-Ganz catheter and transthoracic echocardiography.
Material And Methods:
Prospective study in a Medical/Surgical Intensive Care Unit (ICU). A total of 41 post liver transplant patients were enrolled. CO, CVP, Pw and PASP, were simultaneously determined by two independent observers, utilizing a Swan-Ganz catheter and transthoracic echocardiography, using equations described in the literature. A linear correlation and a Bland-Altman analysis were performed.
Results:
A good correlation between invasive and non- invasive measurements for CO (r=0.97) and CVP (r=0.88) was found, but determinations of Pw (r=0.41) and PASP (r=0.18) did not correlate well. Bias and 95% confidence interval for CO were negligible namely when a CO<6 l/min was considered. Pulsed-wave Doppler-echocardiography underestimates the CO when compared with thermodilution, but the 2 techniques agree on average and can be used interchangeably.
Conclusions:
The non-invasive determination of CO in critical care post liver transplant patients correlates well with the invasive determinations. Transthoracic echocardiography was not appropriate for calculating filling parameters studied. Although the data was obtained in post liver transplant patients, they could be useful in defining the role of echocardiography in the ICU.