Cardiac resynchronization therapy optimization by ultrasonic cardiac output monitoring (USCOM) device
Chung-Wah Siu1, Hung-Fat Tse, Kathy Lee
1Cardiology Division, Department of Medicine, Queen Mary Hospital, The University of Hong Kong, Hong Kong.
Objectives:
We investigated the accuracy and feasibility of a 2D echo-independent ultrasonic continuous wave Doppler cardiac output monitoring device (USCOM) operated by trained nurse for the atrio-ventricular interval (AVI) optimization in cardiac resynchronization therapy (CRT).
Background:
CRT is of proven benefit in patients with advanced chronic heart failure and ventricular conduction delay. Appropriate AVI selection is critical to optimize hemodynamic in CRT. Currently, most non-invasive methods for AVI optimization are often complicated and labor-intensive.
Methods:
USCOM method, Ritter method, and aortic outflow cardiac output method were used to determine the optima AVI in 20 patients with CRT. The accuracy and time for measurement of each method were determined.
Results:
The optimal AVI determined by USCOM method had good correlation with Ritter's method and aortic outflow estimated cardiac output method (r2= 0.78, P < 0.01 and r2= 0.73, P < 0.01, respectively). The optimal AVI determined USCOM method showed good agreement (within 10 msec range) with Ritter's method (85% patients) and aortic outflow estimated cardiac output method (80%). The mean time for determining AVI using USCOM method was shorter than that with aortic outflow method (7.1 +/- 0.7 min vs 12.7 +/- 1.1 min, P < 0.01), whereas the mean time was shortest for Ritter method (4.7 +/- 1.6 min vs 7.1 +/- 0.7 min, P < 0.01).
Conclusion:
USCOM device operated by trained nurse can provide a simple, accurate, and fast non-invasive method for the AVI optimization in CRT population.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Imbalances in Cardiac Output
CHF can occur due to the failure of either side of the heart. Left-side failure leads to pulmonary congestion—the right side continues to send blood...
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac output averages...
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart rate...
Holter Monitor: 24-Hour Monitoring
Cardiac Catheterization IV: Nursing Management