Automatic patient centering for MDCT: effect on radiation dose.

Jianhai Li1, Unni K Udayasankar, Thomas L Toth

  • 1Department of Radiology, Emory University School of Medicine, 1364 Clifton Rd. NE, Atlanta, GA 30322, USA.

AJR. American Journal of Roentgenology
|January 24, 2007
PubMed
Summary
This summary is machine-generated.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Dual-energy CT based scoring to estimate perfusion changes after balloon pulmonary angioplasty in chronic thromboembolic pulmonary hypertension.

JHLT open·2026
Same author

Photon-counting Detector CT Spectral Reconstructions for Radiomics-based Liver Lesion Classification: A Multicenter Study.

Investigative radiology·2026
Same author

Microfluidic automation improves oocyte recovery from follicular fluid of patients undergoing in vitro fertilization.

Nature medicine·2026
Same author

Association between SLCO1B1, apolipoprotein E and ABCG2 genes and lipid response to rosuvastatin: a meta-analysis.

Pharmacogenetics and genomics·2026
Same author

AI-based diagnosis of acute aortic syndrome from noncontrast CT.

Nature medicine·2025
Same author

Are high-dose CT examinations on the rise?

The British journal of radiology·2025

An automatic patient-centering technique significantly reduces radiation dose in multi-detector CT (MDCT) scans. This method optimizes patient positioning, leading to substantial surface dose reductions and improved safety during CT imaging.

Area of Science:

  • Medical Imaging
  • Radiologic Technology
  • Radiation Dose Reduction

Background:

  • Patient positioning accuracy is critical for minimizing radiation exposure in CT scans.
  • Off-center patient placement can lead to increased radiation dose and suboptimal image quality.

Purpose of the Study:

  • To evaluate the impact of an automatic patient-centering technique on radiation dose in MDCT.
  • To quantify dose reduction achieved through optimized patient centering.

Main Methods:

  • Phantom and patient imaging using 64-MDCT with an automatic exposure control technique.
  • Measurement of CT dose index (CTDI) at various off-center distances.
  • Analysis of patient positioning accuracy and dose savings in 63 patients undergoing chest and abdominal CT.

Related Experiment Videos

Main Results:

  • Off-center distances of 30 mm and 60 mm increased peripheral and surface CTDI by 12-18% and 41-49%, respectively.
  • Approximately 95% of patients were not accurately centered in the gantry isocenter.
  • The automatic centering technique achieved a mean radiation dose saving of 13.0%, with a strong correlation between off-center distance and dose reduction.

Conclusions:

  • Improved patient centering within the CT gantry can significantly reduce surface radiation doses.
  • Automatic centering techniques offer a practical solution for optimizing patient positioning and reducing radiation exposure, potentially up to 30% in surface dose.