The Intra-Aortic Balloon Pump
Published on: February 5, 2021
[New method for accurate ECG trigger of intraaortic balloon counterpulsation during off-pump coronary artery bypass
Yosuke Mori1, Takehiko Kaneko, Yasuhiko Aoyama
1Department of Anesthesia, Clinical Engineering, National Hospital Organization, Tokyo Medical Center, Tokyo 152-8902.
Insights
A new BPI 202 interface successfully triggered intra-aortic balloon pumps (IABP) during off-pump coronary artery bypass (OPCAB) surgery. This improved IABP synchronization, ensuring patient safety and stability during cardiac procedures.
Area of Science:
- Cardiovascular Surgery
- Medical Device Technology
- Anesthesiology
Background:
- Off-pump coronary artery bypass (OPCAB) surgery often involves cardiac retraction, which can obscure the R wave on electrocardiograms (ECGs).
- This R-wave interference disrupts intra-aortic balloon pump (IABP) triggering, potentially leading to hemodynamic instability.
- Accurate IABP triggering is crucial for counterpulsation and hemodynamic support during cardiac procedures.
Observation:
- A novel interface, BPI 202, was evaluated for its ability to provide a reliable ECG trigger for IABP during OPCAB.
- The BPI 202 interface processes a simulated R-wave signal derived from an external pacemaker.
- This simulated signal was successfully utilized to synchronize the IABP with cardiac activity.
Findings:
- The BPI 202 interface effectively generated a usable R-wave signal, even with cardiac retraction.
- Successful synchronization between the IABP and the heart was achieved throughout the procedure.
- The interface facilitated intraoperative management in a series of high-risk patients.
Implications:
- The BPI 202 interface demonstrates potential for ensuring synchronous IABP function during OPCAB in high-risk individuals.
- This technology may mitigate the risk of hemodynamic instability associated with IABP trigger loss.
- Wider adoption could enhance the safety and efficacy of OPCAB procedures requiring cardiac manipulation.
Background:
Retractions of the heart required for exposure and construction of distal anastomoses often decrease R-wave amplitude of ECG and interfere with intraaortic balloon pump (IABP) trigger during off-pump coronary artery bypass (OPCAB). Missing R wave trigger results in asynchronous work of IABP and probably produces hemodynamic instability. We report our early experience with a new interface BPI 202 (Osypka Medical, Inc., USA) for sensing accurate ECG trigger for IABP during OPCAB procedure.
Cases:
Six high-risk patients undergoing multivessel OPCAB using BPI 202 are described.
Results:
With the new interface BPI 202, simulated R wave signal could be processed from an external dual chamber pacemaker sensing surface R wave. The simulated R wave was successfully used for controlling IABP and secured a synchronous work between the heart and IABP during heart retraction maneuver. BPI 202, an interface for IABP appeared to facilitate the intraoperative management of our series of patients.
Conclusions:
We believe BPI 202 can produce a synchronous work of IABP during OPCAB procedures to high-risk patients and avoid dangerous hemodynamic instability.