Related Experiment Video
Updated: Jul 17, 2026

Fentanyl Analog Screening using LC-TIMS-TOF MS/MS
Published on: November 8, 2024
Determination of fentanyl in human plasma by head-space solid-phase microextraction and gas chromatography-mass
Habib Bagheri1, Ali Es-haghi, Faezeh Khalilian
1Department of Chemistry, Sharif University of Technology, P.O. Box 11365-9516, Tehran, Iran. bagheri@sharif.edu
Abstract:
A head-space solid-phase microextraction (HS-SPME) method coupled to GC-MS was developed to extract fentanyl from human plasma. The protein binding was reduced by acidification and, eventually, the sample was deproteinized with trichloroacetic acid. The parameters influencing adsorption (extraction time, temperature, pH and salt addition) and desorption (desorption time and temperature) of the analyte on the fibre were investigated and validated for method development. The developed method proved to be rapid, simple, easy and inexpensive and offers high sensitivity and reproducibility. Linear range was obtained from 0.1 ng/ml to 2 microg/ml. The limit of detection was 0.03 ng/ml while an inter-day precision of less than 5% (n=15) could be achieved. The method has been applied for the determination of fentanyl in plasma samples after application of 50 microg/h Duragesic fentanyl patch.
More Related Videos
05:31Gas Chromatography-Mass Spectrometry Paired with Total Vaporization Solid-Phase Microextraction as a Forensic Tool
Published on: May 25, 2021
10:17High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019