Related Experiment Video
Updated: Jul 15, 2026

Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
Cardiac resynchronization therapy: predictive factors of unsuccessful left ventricular lead implant
Alfonso Macías1, Ignacio García-Bolao, Ernesto Díaz-Infante
1Thorax Institute, Hospital Clinic Universitari de Barcelona, Institut d'Investigació Biomèdica August Pi i Sunyer (IDIBAPS), Barcelona, Spain.
Aims:
Cardiac resynchronization therapy is an established therapy for advanced heart failure. However, coronary sinus access and pacing is not achieved in about 5-10% of patients. The aim of this study was to identify predictive factors for failure of left ventricular (LV) lead transvenous implant.
Methods And Results:
We evaluated 212 consecutive patients who received a cardiac resynchronization system. In 26 patients (12.3%), the attempt to pace the LV was unsuccessful. At univariate analysis, in patients with an unsuccessful implant a higher proportion of permanent atrial fibrillation (AF), valvular heart disease, and previous heart surgery were observed. Anteroposterior, longitudinal, and transversal left atrium diameters (LAD) were also larger among patients with an unsuccessful implant. The anteroposterior LAD (APLAD) with an optimal value to predict implant failure was 48.5 mm. At logistic regression analysis, the presence of permanent AF and APLAD were independent predictors of failed implant (OR 7.7, 95% CI 2.5-23.9, P=0.002 and OR 11.7, 95% CI 3.1-37.6, P<0.001, respectively).
Conclusion:
The presence of permanent AF and APLAD are factors that predict unsuccessful pacing from the LV.