[Comparison of plateletpheresis systems]
1Abt. für Transfusionsmedizin und Gerinnungsphysiologie, Uni-Klinikum Marburg.
Summary
Comparing cell separators for platelet collection, the COBE Spectra showed superior separation efficiency and purity. However, no single system was definitively recommended due to varying performance and donor safety considerations.
Area of Science:
- Hematology
- Biomedical Engineering
- Transfusion Medicine
Background:
- Assessing cell separators requires evaluating thrombocyte yield, separation efficiency, and purity of platelet concentrates.
- Comparing separation efficiency across different centers and procedures is challenging.
- Three new-generation cell separators were evaluated: COBE Spectra, Fresenius AS-104, and Fenwal CS-3000 TNX.
Purpose of the Study:
- To compare the performance of three distinct cell separation systems.
- To evaluate thrombocyte yield, separation efficiency, and purity of platelet concentrates.
- To assess potential risks to donors associated with different separation procedures.
Main Methods:
- A multicenter counting study involving twelve centers was referenced.
- Direct comparison of COBE Spectra (n=71), Fresenius AS-104 (n>1100), and Fenwal CS-3000 TNX (n=79) in Marburg.
- Analysis included thrombocyte and leukocyte counts, platelet aggregation, and consideration of anticoagulant citrate dextrose (ACD) blood ratio.
Main Results:
- COBE Spectra demonstrated the highest separation efficiency and lowest leukocyte contamination when the ACD-blood ratio was within -15% of the algorithm.
- AS-104 and CS-3000 TNX showed similar separation efficiencies (approx. 60%), with AS-104 achieving higher purity than CS-3000.
- Reduced ACD ratios in COBE Spectra led to increased platelet aggregation, potentially affecting function and survival; longer donation times were noted.
Conclusions:
- COBE Spectra offers superior separation efficiency and purity, but requires careful management of the ACD-blood ratio to avoid platelet aggregation.
- AS-104 and CS-3000 TNX have comparable separation efficiencies, though AS-104 yields purer platelet concentrates.
- Despite performance differences, a definitive recommendation for a single cell separator system could not be made based on all evaluated criteria.