An Ex Vivo Porcine Model for Hydrodynamic Testing of Experimental Aortic Valve Procedures and Novel Medical Devices
Published on: August 25, 2023
[Accelerated fatigue test in vitro of expanded polytetrafluoroethylene artificial heart valve]
Hai-bo Li1, Wu-jun Wang, Yong Liang
1Department of Cardiothoracic Surgery, Nanfang Hospital, Southern Medical University, Guangzhou 510515, China.
Objective:
To study the durability of expanded polytetrafluoroethylene artificial heart valve (ePTFE AHV).
Methods:
Six ePTFE AHVs were tested for 400 million times against accelerated fatigue using TH-2200 artificial heart valve exosomatic accelerated fatigue instrument. Hydromechanical parameters of fore-and-aft accelerated fatigue test of the 6 AHVs were obtained by TH-1200 artificial heart valve exosomatic pulsatile stream instrument.
Results And Conclusion:
The mean gradient pressure spanning the valve and the effective orific area of ePTFE AHVs did not undergo significant changes after fore-and-aft the fatigue test, but the regurgitation volume and regurgitation rate of ePTFE AHVs were reduced after the accelerated fatigue test, suggesting good durability of ePTFE AHV.